menu
banner

Lab 1200C Vertical Automatic Gas Quenching Tube Furnace with ID 55mm Quartz Tube

Contact Person:Gilia Ding


Email: gilia@inthelaboratory.com

Tel:+86 177 5900 4070
Telephone:+86-177-59004070
Whatsapp: +86 177 5900 4070
Wechat:Dingqiuna



  • Item no.:

    LITH-OTF-1200X-VGQ-6
  • Payment:

    L/C, T/T, Western Union, Credit Cards, Paypal
  • Lead Time:

    7 days
  • Compliance:

    CE Certified
  • Warranty:

    Two years limited standard warranty
  • Product Detail

Lab 1200C Vertical Automatic Gas Quenching Tube Furnace with ID 55mm Quartz Tube


OTF-1200X-VGQ-60 is an automatic gas quenching furnace with an ID 55mm quartz tube and 1200oC max working temperature. It consists of a heating furnace, gas quenching chamber, inert gas injection mechanism. . Automatic procedure included placing a sample into the gas quenching chamber, automatically heating as a program, gas replacement, returning to the gas quenching chamber, and injecting inert gas to quickly cooling.


SPECIFICATIONS

Furnace Structure
  • Double-layer steel structure with air cooling
  • High purity polycrystalline alumina fiber chamber

Power

  • AC 208 - 240V Single Phase, 50/60 Hz
  • 3.5 KW
  • Cable is included without plug

Rated temperature

1100oC

Max. Temperature

1200oC < 30 min

Continuous Temperature

100oC < continuous

Heating Rate

Recommended: <= 10oC /min

Tube Size and Materials

Quartz tube: OD 60 x OD 55 x 800 mm 

Heating Zone

Heating Zone length: 400mm 

Temperature Controllers
  • Eurotherm 3000 Programmable Temperature Controller
  • PID automatic temperature control system
  • 24 programmable process segments
  • DB9 PC communication port
  • +/- 0.1 ºC temperature accuracy
  • All electrical components are approved by UL or MET
  • The pre-configured customized software can automatically record the entire process flow

 Vacuum Level

  • Mechanical Pump (Separate Purchase): < 1e-2 torr 
  • Molecular Pump (Separate Purchase): <1e-5 torr
   

Application Notes

  • Maximum pressure: <0.02Mpa, otherwise, an explosion may happen. 
  • The gas cylinder Must be equipped with a relief valve
  • Use extreme caution to monitor and adjust the pressure (<0.02Mpa) while heating, otherwise, an explosion may happen. 